CASE REPORT  
Year : 2019  |  Volume : 22  |  Issue : 4  |  Page : 435-436
Pseudo inappropriate shock: A technical dilemma


Department of Cardiology, SGPGIMS, Lucknow, Uttar Pradesh, India

Click here for correspondence address and email

Date of Web Publication4-Oct-2019
 

   Abstract 


With liberal availability of high end cardiac implantable devices in recent era, we frequently encounter patients who are recipients of implantable cardioverter defibrillators (ICDs) in our routine clinical practice. Despite improvements in shock detecting algorithms by various manufacturers, incidence of inappropriate shock therapy remains high, it is cause of distress to physicians and patients. Here we present an interesting case of inappropriate shock in one of patient.

Keywords: Implantable cardioverter defibrillator, inappropriate shock, long QT syndrome

How to cite this article:
Shah PJ, Kapoor A. Pseudo inappropriate shock: A technical dilemma. Ann Card Anaesth 2019;22:435-6

How to cite this URL:
Shah PJ, Kapoor A. Pseudo inappropriate shock: A technical dilemma. Ann Card Anaesth [serial online] 2019 [cited 2019 Oct 21];22:435-6. Available from: http://www.annals.in/text.asp?2019/22/4/435/268546





   Introduction Top


Despite proven survival benefits, implantable cardioverter defibrillators' (ICDs) are often associated with inappropriate shocks delivered for causes other than potentially life-threatening ventricular arrhythmias. The reported incidence of inappropriate shocks in patients with ICDs is 4%–25%.[1],[2] Such shocks are not only psychologically disturbing and painful to the patients but also reduce battery longevity and are often proarrhythmogenic.[3],[4] Increased shock delivery and high shock burden associated with recurrent inappropriate shocks is associated with increased mortality.[5]

Automatic QRS morphology template update is an algorithm to help prevent inappropriate therapies in patients with single chamber automatic ICDs in place.[6] We present an interesting case in which an automatic QRS morphology template detection which was inadvertently switched on to auto update” lead to inappropriate ICD therapy and resultant shocks.


   Case Report Top


A 15-year-old boy, with long QT syndrome [Figure 1] and an episode of resuscitated cardiac arrest was implanted with a single chamber ICD (St. Jude, model FORTIFY™ ST VR 1235-40 ICD) in March 2016.
Figure 1: Electrocardiography of the patient showing T-wave inversions and long QT interval (calculated QTc of 499 ms)

Click here to view


He did well for 2 years until he presented with a device shock in February 2018. The device interrogation data revealed multiple episodes of anti-tachycardia pacing and 1 shock therapy of 20 J [Figure 2], all of which were inappropriately delivered in response to high background noise [Figure 3] and [Figure 4].
Figure 2: Interrogation data show multiple episodes of ventricular tachycardia/ventricular fibrillation detected by the device with antitachycardia pacing and a single shock of 20 J

Click here to view
Figure 3: One event of antitachycardia pacing in response to inappropriately sensed background noise

Click here to view
Figure 4: Event of inappropriate shock in response to high background noise inappropriately sensed as ventricular fibrillation

Click here to view


It was found that the automatic QRS morphology template auto-update was turned on. Background noise at the time of recording of reference electrogram for automatic QRS morphology template algorithm detection can lead to inappropriate sensing of background noise as ventricular tachycardia (VT)/ventricular fibrillation resulting in inappropriate therapy. After turning this off, no further inappropriate therapies were delivered [Figure 5].
Figure 5: QRS morphology template auto-update algorithm switched on (left) and off (right)

Click here to view



   Discussion Top


Inappropriate shocks in patients with ICDs can cause anxiety, depression, impaired quality of life, and increased risk of adverse events including higher morbidity and mortality.[7] Important causes of noise/artifact and oversensing causing inappropriate shocks are include external noise interference and lead/connector and muscle noise. Improved discrimination between true and pseudo-arrhythmias is important to reduce the incidence of inappropriate shocks. The QRS morphology template creates an algorithm and stores it for future referencing and matching with tachycardia morphology. Turning on the auto-update can lead to background noise or interference being sensed when the reference electrogram for automatic QRS morphology template algorithm is being recorded periodically by the device. This noise can then be detected by the device as VT or fibrillation leading to inappropriate shocks. In this patients as well, it was noted that the feature was inadvertently programmed on. After turning it off, no further shocks were delivered, and the patient remains asymptomatic on follow-up.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
   References Top

1.
Köbe J, Reinke F, Meyer C, Shin DI, Martens E, Kääb S, et al. Implantation and follow-up of totally subcutaneous versus conventional implantable cardioverter-defibrillators: A multicenter case-control study. Heart Rhythm 2013;10:29-36.  Back to cited text no. 1
    
2.
Jarman JW, Lascelles K, Wong T, Markides V, Clague JR, Till J, et al. Clinical experience of entirely subcutaneous implantable cardioverter-defibrillators in children and adults: Cause for caution. Eur Heart J 2012;33:1351-9.  Back to cited text no. 2
    
3.
Vollmann D, Lüthje L, Vonhof S, Unterberg C. Inappropriate therapy and fatal proarrhythmia by an implantable cardioverter-defibrillator. Heart Rhythm 2005;2:307-9.  Back to cited text no. 3
    
4.
Prudente LA. Phantom shock in a patient with an implantable cardioverter defibrillator: Case report. Am J Crit Care 2003;12:144-6.  Back to cited text no. 4
    
5.
Daubert JP, Zareba W, Cannom DS, McNitt S, Rosero SZ, Wang P, et al. Inappropriate implantable cardioverter-defibrillator shocks in MADIT II: Frequency, mechanisms, predictors, and survival impact. J Am Coll Cardiol 2008;51:1357-65.  Back to cited text no. 5
    
6.
Ross HM, Gerstenfeld EP, Hsia HH, Kocovic DZ. Automatic QRS morphology template update leads to inappropriate therapy inhibition in a single-chamber ICD. Heart Rhythm 2005;2:S293.  Back to cited text no. 6
    
7.
Sears SF Jr., Conti JB. Quality of life and psychological functioning of icd patients. Heart 2002;87:488-93.  Back to cited text no. 7
    

Top
Correspondence Address:
Aditya Kapoor
Department of Cardiology, SGPGIMS, Raebareli Road, Lucknow - 226 014, Uttar Pradesh
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/aca.ACA_109_18

Rights and Permissions


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5]



 

Top