INTERESTING IMAGE  
Year : 2018  |  Volume : 21  |  Issue : 4  |  Page : 442-443
Aortic regurgitation after chemoport catheter insertion: Diagnostic dilemma


1 Department of Cardiac Anaesthesiology, Artemis Hospitals, Gurgaon, Haryana, India
2 Department of Cardiac Surgery, Artemis Hospitals, Gurgaon, Haryana, India

Click here for correspondence address and email

Date of Web Publication17-Oct-2018
 

   Abstract 


Catheter insertion in central vein can potentially cause accidental arterial injury. However it is very unusual to see such catheter in aorta causing aortic regurgitation. Pressure monitoring and Ultrasound guidance can reliably prevent accidental arterial puncture.

Keywords: Aorta, aortic regurgitation, central venous catheter

How to cite this article:
Hanjoora VM, Gupta G, Raut MS. Aortic regurgitation after chemoport catheter insertion: Diagnostic dilemma. Ann Card Anaesth 2018;21:442-3

How to cite this URL:
Hanjoora VM, Gupta G, Raut MS. Aortic regurgitation after chemoport catheter insertion: Diagnostic dilemma. Ann Card Anaesth [serial online] 2018 [cited 2018 Dec 17];21:442-3. Available from: http://www.annals.in/text.asp?2018/21/4/442/243549




A 45-year-old female patient suffering from carcinoma of the breast was advised for chemotherapy. Chemoport catheter was inserted from the right side of the neck at a local hospital. The patient complained of breathlessness on mild exertion along with giddiness few days after the procedure even before the start of chemotherapy. Echocardiographic examination revealed severe aortic regurgitation (AR) with aortic valve endocarditis and was advised for aortic valve replacement [Figure 1]. The patient was referred to our center for further management. Transesophageal echocardiography confirmed (AR) and also showed elongated tubular structure (vegetations?) above aortic valve leaflets [Figure 2]. Chest X-ray showed the course of chemoport catheter more toward the midline [Figure 3]. Hence, computed tomography scan was performed [Figure 4].
Figure 1: Transthoracic echocardiography showing aortic regurgitation

Click here to view
Figure 2: Transesophageal echocardiography aortic valve short-axis view showing catheter

Click here to view
Figure 3: Chest X-ray showing the course of the catheter

Click here to view
Figure 4: Computed tomography scan images showing chemoport catheter in (a) Brachiocephalic artery, (b) Ascending aorta, and (c) Aortic root

Click here to view


The passage of chemoport catheter was observed through the right internal carotid artery and ascending aorta till aortic valve. The “arterial” catheter was removed and internal carotid artery was repaired. Subsequently, chemoport catheter was inserted through the left internal jugular vein under ultrasound guidance, and proper placement was confirmed by pressure monitoring and chest X-ray. Now, the echocardiogram revealed absolutely no AR.

Incidences of accidental arterial puncture occur in about 1% and 2.7% of jugular and subclavian approaches, respectively.[1] Arterial cannulation was reported in <1% of cases.[1] However, inadvertent carotid or subclavian arterial injury can lead to serious complications such as pseudoaneurysm, hematoma formation, stroke, and fistula.[2] Direct aortic cannulation is an exceedingly uncommon event with significant morbidity and mortality.[1]

Arterial puncture can be recognized by pulsatile flow, but these findings may get obscured in patients with hypoxia and shock. Blood gas analysis may be helpful to confirm the arterial blood, but it is not suitable during the procedure due to time delay. Chest X-ray can guide about the course of the catheter, but it is not always a reliable method for preventing or readily detecting inadvertent arterial cannulation.

Pressure monitoring and ultrasound guidance can reliably prevent accidental arterial puncture.[3] Ultrasound guidance, in comparison with anatomical landmark techniques, was associated with 71% decrease in catheter-related complications for internal jugular vein placement.[4]

A rare case of iatrogenic type A aortic dissection caused during mediport catheter and venous catheter insertion into the false lumen of Type B aortic dissection has been reported.[5],[6] However, in the present case, apart from arterial cannulation, the catheter went up to the aortic valve causing AR by interfering with the aortic valve coaptation. In the absence of any previous medical records, we presume that chemoport catheter insertion might have been performed using anatomical landmark technique and not by ultrasound guidance at previous hospital. A chest radiograph should have been obtained immediately after the catheter was inserted to verify its correct position. Failure to obtain a chest radiograph under these circumstances does not meet the minimum standard of care. Even though echocardiographic finding of catheter in aortic root was misinterpreted as vegetations like structure at the previous hospital, computed tomography scan imaging definitely clarified the diagnostic dilemma in this unfortunate clinical situation.

Declaration of patient consent

The authors certify that they have obtained all appropriate patient consent forms. In the form the patient(s) has/have given his/her/their consent for his/her/their images and other clinical information to be reported in the journal. The patients understand that their names and initials will not be published and due efforts will be made to conceal their identity, but anonymity cannot be guaranteed.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
   References Top

1.
Blaivas M. Video analysis of accidental arterial cannulation with dynamic ultrasound guidance for central venous access. J Ultrasound Med 2009;28:1239-44.  Back to cited text no. 1
    
2.
Garg N, Noheria A, McPhail IR, Ricotta JJ 2nd. Embolic strokes after peripherally inserted central catheter placement. Ann Vasc Surg 2010;24:1133.e1-4.  Back to cited text no. 2
    
3.
Othman H, Rosman H, Saravolatz L, Rama K. Aortic cannulation: An unexpected complication of central venous catheter placement. J Cardiothorac Vasc Anesth 2017;31:e52-3.  Back to cited text no. 3
    
4.
Rezayat T, Stowell JR, Kendall JL, Turner E, Fox JC, Barjaktarevic I, et al. Ultrasound-guided cannulation: Time to bring subclavian central lines back. West J Emerg Med 2016;17:216-21.  Back to cited text no. 4
    
5.
Tsukashita M, Deanda A, Balsam L. Type A aortic dissection: A rare complication of central venous catheter placement. J Card Surg 2014;29:368-70.  Back to cited text no. 5
    
6.
Bas A, Goksedef D, Kandemirli SG, Gulsen F, Numan F. Central venous catheter insertion into the false lumen of a complicated aortic dissection. Scott Med J 2017;62:115-8.  Back to cited text no. 6
    

Top
Correspondence Address:
Monish S Raut
Department of Cardiac Anaesthesiology, Artemis Hospitals, Gurgaon, Haryana
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/aca.ACA_79_18

Rights and Permissions


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4]



 

Top