LETTER TO EDITOR  
Year : 2011  |  Volume : 14  |  Issue : 2  |  Page : 154-155
A simple method of electrocardiogram: Controlled central venous catheterization


Department of Anesthesia and Critical Care,Christian Medical College, Vellore, India

Click here for correspondence address and email

Date of Web Publication25-May-2011
 

How to cite this article:
Raviraj R, Korula G, Subramani K, Shalinicynthia S. A simple method of electrocardiogram: Controlled central venous catheterization. Ann Card Anaesth 2011;14:154-5

How to cite this URL:
Raviraj R, Korula G, Subramani K, Shalinicynthia S. A simple method of electrocardiogram: Controlled central venous catheterization. Ann Card Anaesth [serial online] 2011 [cited 2019 Jul 16];14:154-5. Available from: http://www.annals.in/text.asp?2011/14/2/154/81574


The Editor,

Central venous catheter (CVC) placement is an important intervention for fluid management, long-term antibiotic therapy, parenteral nutrition, and chemotherapy. The common method of assessing the correct placement of CVC is by the use of radiographs, which is carried out post the procedure, with the possibility of more attempts if unsatisfactory. Fluoroscopy may not be feasible in all the areas where CVC cannulation has been performed. Other methods described in recent times need expensive equipment such as ultrasound or echocardiography. Although electrocardiogram (ECG) guidance for CVC positioning was suggested as early as 1949, [1] this method is not widely used. A recent studyby Von Hellerstein [2] has shown that the chance of correct placement of CVC using the ECG-controlled technique is 96% versus 76% with the conventional technique. Although, monitoring the P-wave morphology of the ECG by connecting the guide wire with the universal adapter is an established technique to position the central line at the correct depth, [3],[4] only a few central line kits (B Braun) are packed with the connecting cable to the guide wire. Also the unavailability of a universal adapter (which connects the guide wire and the ECG lead) limits the usage of this technique. We describe below a simple alternate technique to overcome this situation, without any extra cost, with instruments that are available in all the wards.

Prior to cannulation, a guide wire is passed into the cannula and the point where the tip corresponds with the tip of the cannula is measured. After skin dilatation, the CVC is threaded over the guide wire and adjusted up to the measured point. The guide wire is clamped close to the proximal end of CVC (measured point). The right ECG lead is attached to the guide wire by using the available clamps as shown in the [Figure 1]. This helps to complete the electrical circuit from the guide wire to the ECG lead. When the ECG monitor is set for lead II, a biphasic P wave corresponding to the guide wire tip is positioned at the mid right atrium, and withdrawal of the guide wire tip to the superior vena cava-right atrial (SVC-RA) junction produces a tall P wave, which equals the R wave height. Further withdrawal of the guide wire into the SVC reduces the P wave height. These changes in P wave morphology are observed and anchored where the morphology changes from the biphasic to the tall peak pattern if an SVC-RA junction is the site one is looking for, or where the size of the P wave becomes less than R for the superior vena cava. Two clamps can be used to maintain sterility during the procedure. By this simple alternate technique, proper placement of the central line at an appropriate depth can be achieved. The newer ECG monitors have an in-built safety mechanism to prevent electrical microshock; the output of an ECG monitor's power supply is electrically isolated from the patient by placing a very high impedance between the monitor and the patient's ECG leads. [5] Before using this technique, one has to make sure that this safety feature is present in the monitor.
Figure 1: ECG-controlled technique with alternate method

Click here to view


 
   References Top

1.von Hellerstein HK, Pritchard WH, Lewis RL. Recording of intracavity potentials through a single-lumen saline filled cardiac catheter. Proc Soc Exp Biol Med 1949;71:58-60.  Back to cited text no. 1
    
2.Gebhard RE, Szmuk P, Pivalizza EG, Melnikov V, Vogt C, Warters RD. The accuracy of electrocardiogram-controlled central line placement. Anesth Analg. 2007;104:65-70.  Back to cited text no. 2
    
3.Joshi AM, Bhosale GP, Parikh GP, Shah VR. Optimal positioning of right-sided internal jugular venous catheters: Comparison of intra-atrial electrocardiography versus Peres' formula. Indian J Crit Care Med 2008;12:10-4.  Back to cited text no. 3
[PUBMED]  Medknow Journal  
4.Corsten SA, van Dijk B, Bakker NC, de Lange JJ, Scheffer GJ. Central venous catheter placement using the ECG-guided Cavafix-Certodyn SD catheter. J Clin Anesth 1994;6:469 -72.  Back to cited text no. 4
[PUBMED]    
5.Ehrenwerth J, Sefert HA. Electrical and fire safety. In: Barash PG, Cullen BF, Stoelting RK, editors. Clinical anesthesia. 5 th ed. New york: Lippincott Williams and Witkins; 2006.p.162-3.  Back to cited text no. 5
    

Top
Correspondence Address:
R Raviraj
Assistant Professor in Anesthesiology,Christian Medical College and Hospital, Vellore
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0971-9784.81574

Rights and Permissions


    Figures

  [Figure 1]



 

Top